范圍:
可實(shí)現(xiàn)實(shí)時(shí)顯示、控溫、測(cè)量等功能。主要用于非金屬材料(如塑料、橡膠、尼龍、電絕緣材料等)的熱變形及維卡軟化點(diǎn)測(cè)定,是化工企業(yè)、科研單位、大專院校等行的理想測(cè)試工具。
參數(shù):
1、最大變形誤差:?0.01毫米
2、升溫速率:120?5℃/h(12?1℃/6min)或50?3℃/h(5?0.5℃/6min)
3、冷卻方法:150℃以上自然冷卻,150℃以下自然冷卻或水冷
4、形變測(cè)量方法:千分表5、變形測(cè)量范圍:-0.1mm——5.0mm
6、負(fù)載桿和壓頭總質(zhì)量:88g
7、試樣架數(shù):三架一溫
8、加 熱 介 質(zhì):甲基硅油(需方自備)
9、加 熱 功 率:2.5KW
10、溫度分辨率:0.01℃ 3、最大溫度誤差:?0.5℃
11、具有上限溫度任意設(shè)定(環(huán)境溫度——300℃)的保護(hù)功能或最高上限300℃的保護(hù)
12、溫度控制范圍:環(huán)境溫度——300℃
13、試驗(yàn)最大負(fù)荷:
A:維卡試驗(yàn)負(fù)荷:GA=10N?0.2N(即1000g)、 GB=50N?1N(即5000g)
B:熱變形試驗(yàn)負(fù)荷:F=2?δ?b?h?h/(3?L) 式中:F——試樣所需加的載荷,單位N δ——熱變形標(biāo)準(zhǔn)彎曲正應(yīng)力:
方法
A:使用纖維應(yīng)力1.80MPa 方法
B:使用纖維應(yīng)力0.45MPa
方法C:使用纖維應(yīng)力8.00MPa b——試樣的寬度,單位m h——試樣的高度,單位m L——兩支點(diǎn)間的距離,0.100m 14、電 源:AC220V?10%、50Hz、30A 15、需方自備甲基硅油:閃點(diǎn)300度、粘度200厘斯,20升。
主要配置:
1.試驗(yàn)主機(jī)一臺(tái)
2.溫度傳感器一只
3.光柵千分表三塊
4.熱變形及維卡測(cè)頭各三套
5. 打印機(jī)一臺(tái)
6. 專用工具一套
特點(diǎn):
本儀器符合ISO75-1974,ISO306-1974,GB1633-79,GB1634-79標(biāo)準(zhǔn)的要求,可實(shí)現(xiàn)實(shí)時(shí)顯示、控溫、測(cè)量等功能。
安裝與操作說(shuō)明:
試驗(yàn)前的準(zhǔn)備工作。根據(jù)試驗(yàn)類型選擇試驗(yàn)壓頭,熱變形為圓角壓頭,維卡為針型壓頭。將所需的壓頭與負(fù)載桿固定好。根據(jù)試驗(yàn)類型和試樣的尺寸計(jì)算載荷:a、熱變形試驗(yàn):根據(jù)試樣的尺寸,選取載荷b、維卡試驗(yàn):只有兩種規(guī)定的負(fù)荷A法9.81N 即1000g B法 49.05N即5000g 砝碼的選配見(jiàn)(9)砝碼質(zhì)量表 打開(kāi)攪拌電機(jī)電源,指示燈亮,攪拌電機(jī)工作,攪拌速度可以通過(guò)攪拌電機(jī)控制盒上的旋鈕調(diào)節(jié)(溫度高時(shí)用慢速,溫度低時(shí)用快速),按照以上指定的程序檢查電氣,無(wú)誤后繼續(xù)做下面的工作。 安裝試樣 將試樣架抬(升)出油面,將熱變形試樣放在支撐架上(維卡試樣放在平面上),放下負(fù)載桿,將試樣壓緊。 將試樣架放(降)回油池內(nèi),將相應(yīng)質(zhì)量的砝碼放在托盤(pán)上。要求砝碼放正。 將千分表固定在砝碼上方,使千分表內(nèi)芯懸于砝碼之上,內(nèi)芯上端有2-5mm的空間。 升溫速率設(shè)定:共有兩檔升溫速率,120℃/h,50℃/h。每次試驗(yàn)時(shí)需要對(duì)千分表進(jìn)行調(diào)零,方可進(jìn)行試驗(yàn),調(diào)零時(shí)輕輕旋動(dòng)旋鈕,使千分表置于零區(qū),預(yù)置3-5分鐘。 啟動(dòng)工作:當(dāng)一切準(zhǔn)備工作就緒后,試驗(yàn)開(kāi)始。 試驗(yàn)結(jié)束,記錄下溫度;移開(kāi)千分表及砝碼,抬(升)起試驗(yàn)架,取下試樣。注意不要把試樣誤入油池。 降溫采用氣冷或水冷,也可自然冷卻。冷卻時(shí),可抬起或?qū)⒃嚇蛹軡撊胗统亍?關(guān)閉主機(jī)電源、攪拌電機(jī)電源。
Installation and operation instructions:
Preparation work before the experiment. Select the test pressure head according to the test type, with rounded pressure head for thermal deformation and needle pressure head for Vicat. Secure the required pressure head with the load rod. Calculate the load based on the type of test and the size of the sample: a. Thermal deformation test: Select the load based on the size of the sample. b. Vicat test: There are only two specified loads: A method 9.81N, which is 1000g, B method 49.05N, which is 5000g. The selection of weights is shown in (9) Weight quality table. Turn on the power supply of the mixing motor, the indicator light is on, and the mixing motor works, The mixing speed can be adjusted through the knob on the mixing motor control box (use slow speed when the temperature is high, and use fast speed when the temperature is low). Check the electrical system according to the specified procedure above, and continue with the following work once there are no errors. Install the sample and lift the sample holder out of the oil surface. Place the thermal deformation sample on the support bracket (Vicat sample on a flat surface), lower the load rod, and press the sample tightly. Place (lower) the sample holder back into the oil tank and place the corresponding weight on the tray. Require the weights to be placed upright. Fix the dial gauge above the weight, so that the inner core of the dial gauge is suspended above the weight, with a space of 2-5mm at the upper end of the inner core. Heating rate setting: There are two heating rates, 120 ℃/h and 50 ℃/h. During each test, the dial indicator needs to be zeroed before the test can be conducted. Gently rotate the knob to set the dial indicator in the zero zone and preset for 3-5 minutes. Start work: When all preparations are ready, the experiment begins. After the experiment, record the temperature; Remove the dial gauge and weight, lift (lift) the test stand, and remove the sample. Be careful not to accidentally drop the sample into the oil pool. Air cooling or water cooling or Free cooling is adopted for cooling. When cooling, the sample holder can be lifted or submerged into the oil pool. Turn off the power supply of the host and mixing motor.
特點(diǎn)及用途:
適用于測(cè)試高分子材料的維卡軟化點(diǎn)溫度和熱變形溫度,作為控制質(zhì)量和鑒定新品種熱性能的一個(gè)指標(biāo),由百分表測(cè)量形變,溫控儀設(shè)定升溫速度,試樣架自動(dòng)升降,一次可試驗(yàn)三個(gè)試樣。操作方便、設(shè)計(jì)新穎、外形美觀、可靠性高。符合GB/T 1633《熱塑 性塑料軟化溫度(VST)的測(cè)定》、GB/T1634《塑料彎曲負(fù)載熱變形溫度 試驗(yàn)方法》、GB 8802《硬聚氯乙烯(PVC-U)管材及管件維卡軟化溫度測(cè)定方法》以及ISO75、ISO306、ISO2507等標(biāo)準(zhǔn)要求。
售后管理:
我公司實(shí)現(xiàn)計(jì)算機(jī)化管理,實(shí)行客戶定期電話回訪制度,定期復(fù)查設(shè)備的工作情況,定期電話指導(dǎo)用戶對(duì)設(shè)備進(jìn)行保養(yǎng)和檢測(cè),以便設(shè)備正常運(yùn)轉(zhuǎn),跟蹤客戶的設(shè)備使用情況,以便及時(shí)對(duì)設(shè)備進(jìn)行維護(hù)。及時(shí)提供貨源,確保全國(guó)范圍內(nèi)2-10內(nèi)到貨,持續(xù)穩(wěn)定地為代理商做全面服務(wù)工作,提供合理的運(yùn)作空間。公司有專業(yè)的技術(shù)人員為您安裝指導(dǎo)。
儀器包裝內(nèi)附有說(shuō)明書(shū),專業(yè)培訓(xùn)人員為客戶指定操作者講解儀器正常操作流程、操作注意事項(xiàng)及儀器的日常維護(hù)要求;也可以根據(jù)客戶要求提供上門(mén)培訓(xùn)服務(wù)。一年保修,終身維護(hù)!
歡迎您選用我公司開(kāi)發(fā)研制的各類檢測(cè)儀器。在使用過(guò)程中,如出現(xiàn)什么問(wèn)題,可隨時(shí)和我們聯(lián)系,您將得到良好的售后服務(wù)。真誠(chéng)的希望您能將寶貴的意見(jiàn),豐富的經(jīng)驗(yàn)和獨(dú)到的見(jiàn)解反饋給我們!
業(yè)務(wù)咨詢:932174181 媒體合作:2279387437 24小時(shí)服務(wù)熱線:15136468001 盤(pán)古機(jī)械網(wǎng) - 全面、科學(xué)的機(jī)械行業(yè)免費(fèi)發(fā)布信息網(wǎng)站 Copyright 2017 PGJXO.COM 豫ICP備12019803號(hào)